may take a moment, but there will be the sound of x-ray being fired and afterwards,
there will be another User Instruction to release the Scan button.
During this calibration, an additional window will open up and be populated with
numbers and lights, there is nothing to do with this window, just allow it to go through
its process – do not attempt to close it.
After each time that x-ray is fired, the panel will need to discharge. There is a
countdown box that will display the seconds until the panel is ready and then the
User Instructions will be updated as to what to do next. X-ray will be fired THREE
times throughout this calibration.
WARNING
The X-ray device may be dangerous to the
Patient and Operator if you do not observe and follow operating
instructions. Do not operate this system unless you have
received training to perform a procedure
WARNING
The X-ray device may be dangerous to the
Patient and Operator if you do not observe and follow operating
instructions. Do not operate this system unless you have
received training to perform a procedure
Содержание pedCAT
Страница 32: ...31 0473 90001 Rev 07 26 13 Rear Connector Panel Voltage Nameplate location System Label location ...
Страница 34: ...33 0473 90001 Rev 07 26 13 Indicator Panel on machine Patient Platform label Patient Seat label ...
Страница 35: ...34 0473 90001 Rev 07 26 13 Handle Bar Label Cleaning Instructions Label Pinch Point Label Laser Label ...
Страница 37: ...36 0473 90001 Rev 07 26 13 Tube Head Label Beam Limiter Label X Ray Power Supply label ...
Страница 40: ...39 0473 90001 Rev 07 26 13 Laser Lights Button ...
Страница 93: ...92 0473 90001 Rev 07 26 13 Beam Path and Angulation ...
Страница 97: ...96 0473 90001 Rev 07 26 13 ...
Страница 144: ...143 0473 90001 Rev 07 26 13 ...
Страница 168: ...167 0473 90001 Rev 07 26 13 ...