POWER
OFF/STANDBY
ON
2.0
Physical Description
Systolic Display
Displays systolic pressure.
Mean Display
Displays mean arterial pressure.
Also displays the pressure of the
cuff during measurement.
Diastolic Display
Displays diastolic pressure
Pulse Rate Display
Displays pulse rate (beats per
minute) upon completion of mea-
surement.
Systolic Alarm Display
Displays systolic High and Low
alarm limits.
Pulse Level Indicators
Displays a red signal proportional to
the amplitude of the oscillation the
monitor senses during cuff defla-
tion.
Battery Indicator
Lights in green during battery
operation; flashes in red when
battery is low.
Power Switch
Controls the AC or selected
battery power to the monitor
Start/Stop Switch
Starts initial cuff inflation. If pushed
during inflation, it automatically
stops and deflates the cuff.
Interval Switch
Selects either "NO INTERVAL" or
"CONTINUOUS" Interval for
consecutive blood pressure monitor-
ing.
Alarms - SET, HIGH, LOW,
MUTE
Set - Sets LCD in alarm mode
High - Sets high alarm limit
Low - Sets low alarm limit
Mute - Silences audible alarm
List
Scrolls up or down to read previous
data. Pressing the scroll down
switch for more than three (3)
seconds will enable "ALL DATA"
print mode. Pressing both the up
and down switches for three (3)
seconds will erase all stored data.
Paper feed
Feeds the printer paper
Clock/Elapsed Time Display
Alternately displays 24 hour clock
and elapsed time meter.
ADULT
NEONATAL
2.1
Front Panel Functions
LCD Screen
Displays data for previous four (4)
measurements; maximum of 400 lines
of memory can be scrolled. Alarm
limits and error codes are also
displayed on this screen.
Adult/Neonatal
Displays which mode the monitor is
in, adult/pediatric or neonatal.
2-1
START
STOP
SET
HI
LO
PAPER
FEED
LIST
CLOCK SYSTOLIC DIASTOLIC PULSE
(mmhg)
(mmhg)
(bpm)
07:00 120 75 68
10:30 122 71 68
122
SYSTOLIC
mmhg
87
MEAN
mmhg
71
DIASTOLIC
mmhg
68
PULSE
Bpm
150
SYS HIGH
70
SYS LOW
10:30
CLOCK/ELAPSED
INTERVAL
oFF
PULSE LEVEL