8-3
•
E01 and E02 RAM/ROM Failure- These messages alert the user to failure of the self test of the unit.
If
one of these messages appear, restart the unit and if the message reappears the unit must be serviced
by the appropriate personnel
.
•
E03
Cuff Disconnect
-
Is displayed when the monitor did not inflate to a pressure above 10 mmHg within 10
seconds.
Verify cuff and hose for extreme air leakage. Also verify that the connections of the above
to the monitor or cuff are made and tight. If the pump did not even start, the monitor requires repair
.
•
C11 Check Cuff- Is displayed when the monitor did not inflate to the inflation pressure setting within 30
seconds.
Verify for a small leak in the monitor, cuff, hose, or connection.
•
C12 Check Cuff- The cuff deflated to a pressure of 10 mmHg, and no measurement was z
that a neonatal
cuff is not being used in the adult mode.