![Ceyear 1465 Series Скачать руководство пользователя страница 91](http://html1.mh-extra.com/html/ceyear/1465-series/1465-series_programming-manual_2584913091.webp)
3 SCPI
3.3 Instrument subsystem command
84
Setting format:
[:SOURce]:DM:IQADjustment:QSKew <val>
Query format:
[:SOURce]:DM:IQADjustment:QSKew?
Parameter:
<Offset> I/Q adjustment orthority offset.
Range: 0deg [-10.00deg
,
+10.00deg].
Example:
:DM:IQADjustment:QSKew 30deg
Set the Orthority Offset for I/Q adjustment to 30deg.
Reset status:
0deg
Key Entry:
【
I/Q
】
—>[ I/Q Input Adj]—>[Orthority Offset]
[:SOURce]:DM:IQADjustment[:STATe] <State>
Function:
This command enables I/Q Adjust ON/OFF. After this function is turned on, the
parameters adjusted: Gain Balance, I Offset, Q Offset and Orthority Offset are
superimposed on adjustment circuit; after this function is turned off, the above
parameter values are not used, but the modulation attenuation is not affected by the
ON/OFF
state
of
I/Q
Adjust.
For
related
commands,
see
―
:DM:MODulation:ATTenuation
‖,―
:DM:IQADjustment:EXTernal:IQATten
‖.
Setting format:
[:SOURce]:DM:IQADjustment[:STATe] ON|OFF|1|0
Query format:
[:SOURce]:DM:IQADjustment[:STATe] ?
Parameter:
<State> Boolean data is as follows:
ON | 1: I/Q Adjust is turned on,
OFF | 0: I/Q Adjust is turned off.
Example:
:DM:IQADjustment 1 Turn on I/Q Adjust.
Reset status:
0
Key Entry:
【
I/Q
】
—>[ I/Q Input Adj]—>[ I/Q Adjust ON OFF]
[:SOURce]:DM:MODulation:ATTenuation <Atten>
Function:
This command sets the attenuation of I/Q signal modulated by the signal generator RF
channel. The output of the output attenuation can be set when Attenuation is in
Manual state. Attenuation is valid even if I/Q Adjust is turned off. For related
commands, see ―
:DM:MODulation:ATTenuation:AUTO
‖,―
:DM:IQADjustment[:STATe]
‖.
Setting format:
[:SOURce]:DM:MODulation:ATTenuation <val>
Query format:
[:SOURce]:DM:MODulation:ATTenuation?
Parameter:
<Atten> I/Q Modulation Attenuation.
Range: 12.00dB[0.00dB
,
40.00dB].
Example:
:DM:MODulation:ATTenuation 10dB Attenuation of I/Q modulator is 10dB.
Reset status:
0dB
Key Entry:
【
I/Q
】
—>[Attenuation]—>[Attenuation]