2
Remote Control
2.1 Remote control basis
12
-7.89E-01
exponent marker e may be upper or lower case
+256
l allowed
5
decimal point may be leading
Extended numeric parameter
Most measurements related to Instrument-specific commands use extended numeric parameters to
specify the physical quantities. Extended numeric parameters receive all numeric parameters and
additional special values. All extended numeric parameters receive MAXimum and MINimum as
parameter values
。
Other special values, such as: UP and DOWN are received by the instrument
parsing capability. SCPI table will list all valid parameters.
Note: Extended numeric parameters are not applicable to common commands or STATus subsystem
commands.
Examples of extended numeric parameters:
101
Numeric parameter
1.2GHz
GHz may be used an exponent (E009)
200MHz
MHz may be used an exponent (E006)
-100mV
-100 millivolt
10DEG
10 degrees
MAXimum
maximum valid setting
MINimum
minimum valid setting
UP
increase a step
DOWN
decrease a step
Discrete parameter
When there are a finite number of parameter values to be set, discrete parameters are used for
identification. A discrete parameter uses mnemonics to represent each valid setting. Like the SCPI
mnemonics, the discrete parameter mnemonics can be set in long and short formats, with both
capitalized and lowercase characters.
The following example illustrates the combined use of discrete parameter and command.
:TRIGger[:SEQuence]:SOURce BUS|IMMediate|EXTernal
BUS
GPIB, LAN, RS-232 trigger
IMMediate
Immediate trigger
EXTernal
External trigger
Boolean parameter
Boolean parameters represent a single binary condition that is either true or false. There are only four
possible representations for a Boolean parameter.
Samples of Boolean parameters:
ON
Logical true
OFF
Logical false
1
Logical true
0
Logical false
String parameter
A string parameter allows the ASCII string to be sent as a parameter. Single and double quotes are used
as separators.