3 SCPI
3.3 Instrument subsystem command
49
Key Entry:
【
Sweep
】
—>[List Sweep]—>[Insert Counts]
[:SOURce]:LIST:FILL:STARt <FreqStart>
Function:
This command sets Freq Start of List Sweep, which is used with Freq Stop and Insert
Counts to automatically generate list sweep points. For setting the stop frequency and
sweep
points
of
the
list,
see
the
commands
―:LIST:FILL:STOP‖,―:LIST:FILL:POINts‖,:LIST:FILLL:POWer‖,―LIST:FILL:DWELl‖.
Setting format:
[:SOURce]:LIST:FILL:STARt <val>
Query format:
[:SOURce]:LIST:FILL:STARt?
Parameter:
<FreqStart> Freq Start of List Sweep.
Model Range
1465A [100kHz~3GHz]
1465B [100kHz~6GHz]
1465C [100kHz~10GHz]
1465D [100kHz~20GHz]
1465E [100kHz~40GHz]
1465H [100kHz~50GHz]
1465L [100kHz~67GHz]
Example:
:LIST:FILL:STARt 300MHz Set Freq Start of List Sweep to 300 MHz.
Key Entry:
【
Sweep
】
—>[List Sweep]-> [Freq Start]
[:SOURce]:LIST:FILL:STOP <FreqStop>
Function:
This command is used for setting the stop frequency for list sweep. It is used together
with the start frequency and number of points of the list to generate the list sweep
point automatically. For setting start frequency and number of sweep points of the list,
see
the
commands
―
:LIST:FILL:STARt
‖,―
:LIST:FILL:POINts
‖,―
:LIST:FILLL:POWer
‖,―
LIST:FILL:DWELl
‖.
Setting format:
[:SOURce]:LIST:FILL:STOP <val>
Query format:
[:SOURce]:LIST:FILL:STOP?
Parameter:
<FreqStop> Freq Stop of List Sweep.
Model
Range
1465A [100kHz~3GHz]
1465B [100kHz~6GHz]
1465C [100kHz~10GHz]
1465D [100kHz~20GHz]
1465E [100kHz~40GHz]
1465H [100kHz~50GHz]
1465L [100kHz~67GHz]
Example:
:LIST:FILL:STOP 1GHz Set Freq Stop of List Sweep to 1 GHz.
Key Entry:
【
Sweep
】
—>[List Sweep]-> [Freq Stop]