![Ceyear 1465 Series Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/ceyear/1465-series/1465-series_programming-manual_2584913051.webp)
3 SCPI
3.3 Instrument subsystem command
44
Parameter:
<State> Boolean data is as follows:
ON | 1: Attenuation is AUTO,
OFF | 0: Attenuation is Manual.
Example:
:POWer:ATTenuation:AUTO 0 Attenuation Style is Manual.
Reset status:
1
Key Entry:
【
Amplitude
】
—>[Attenuation Config]—>[Attenuation Style Auto Manual]
[:SOURce]:POWer[:LEVel][:IMMediate][:AMPLitude] <Ampl>
Function:
This command sets the output amplitude level of the signal generator.
Setting format:
[:SOURce]:POWer[:LEVel][:IMMediate][:AMPLitude] <value>
Query format:
[:SOURce]:POWer[:LEVel][:IMMediate][:AMPLitude]?
Parameter:
<Ampl> Amplitude level value.
Range: -135dBm [-135dBm
,
+30dBm].
Example:
:POWer 0dBm The output amplitude level is 0 dBm.
Reset status:
-115dBm
Key Entry:
【
Amplitude
】
[:SOURce]:POWer[:LEVel][:IMMediate]:OFFSet <PowOffset>
Function:
The command is the actual output amplitude offset of the signal generator. When the
value is non-
zero, an offset indicator ―*‖ is displayed in amplitude display area and the
displayed amplitude value is the actual amplitude power plus the amplitude offset.
Such amplitude offset does not change the actual output amplitude of the signal
generator. Only the displayed amplitude value is changed.
Setting format:
[:SOURce]:POWer[:LEVel][:IMMediate]:OFFSet <value>
Query format:
[:SOURce]:POWer[:LEVel][:IMMediate]:OFFSet?
Parameter:
<PowOffset> Ampl Offset.
Range: 0dB [-100dB
,
+100dB].
Example:
:POWer:OFFS -10dB Ampl Offset is -10 dB.
Reset status:
0dB
Key Entry:
【
Amplitude
】
—>[Basic Config]->[Ampl Offset]
[:SOURce]:POWer:REFerence <PowRef>
Function:
Ampl Ref can be set when Ampl Ref Switch is turned on. For the state of power
reference switch, see the command ―
:POWer:REFerence:STATe
‖. When Ampl Ref
Switch is turned on, an indicator ―*‖ is displayed in amplitude display area and the
displayed amplitude value = actual output amplitude - amplitude reference value.
For example: When current CW output power is 1dBm, if the power reference is set to
1dBm, the displayed CW output power is based on this power reference. Therefore,
the power display area shows 0dBm, and the actual output frequency of the signal