3 SCPI
3.3 Instrument subsystem command
64
Note:
for setting only.
[:SOURce]:PULM:INTernal:PTRain:POINts?
Function:
This command is used for querying points of current pulse train. "Zero" indicates that the
current list is empty, and "Nonzero" indicates the actual points in the current list. It
should be noted that, if the user operates the interface manually, the index in the pulse
trains list will start from 0; that is, if the queried number of points in the list is 1, the
index actually shown in the list is 0.
Query format:
[:SOURce]:PULM:INTernal:PTRain:POINts?
Return value:
<Num> Integer data; returned points of pulse trains list.
Range: 0[0
,
1023].
Example:
[:SOURce]:PULM:INTernal:PTRain:POINts? Query points of current pulse trains list.
Reset status:
0
Key Entry:
None
Note:
for query only.
[:SOURce]:PULM:INTernal:PTRain:PRESet
Function:
This command is used for deleting all points in the pulse trains list. If the current list is
empty, no operation will be conducted. The user may query the points in the current
list before deletion operation. For the command related to pulse train points, see
―
PULM:INTernal:PTRain:POINts
‖.
Setting format:
[:SOURce]:PULM:INTernal:PTRain:PRESet
Example:
[:SOURce]:PULM:INTernal:PTRain:PRESet Delete all point in the pulse train list.
Key Entry:
【
PULSE
】
—>[Pulse Train]—>[Edit Pulse Train] —>[Del All]
Note:
for setting only.
[:SOURce]:PULM:INTernal:SLIDing:STEP <StepTime>
Function:
When the pulse source is set to Sliding, this command is used for setting pulse sliding
step. Based on the current pulse period, the pulse signal will increase progressively
with the set sliding step. For example, when the step point is set to 1024, current pulse
period is 1ms and sliding step is 100us, the pulse period will change between 1ms and
103.4ms.
Setting format:
[:SOURce]:PULM:INTernal:SLIDing:STEP <val><time unit>
Query format:
[:SOURce]:PULM:INTernal:SLIDing:STEP?
Parameter:
<StepTime> Sliding step.
Range: 100ns[0s
,
42ms].
Example:
[:SOURce]:PULM:INTernal:SLIDing:STEP 100us Set pulse sliding step to 100us.
Reset:
100ns
Key Entry:
【
PULSE
】
—>[Sliding]—>[Sliding Step]