![Ceyear 1465 Series Скачать руководство пользователя страница 109](http://html1.mh-extra.com/html/ceyear/1465-series/1465-series_programming-manual_2584913109.webp)
3 SCPI
3.3 Instrument subsystem command
102
Parameter:
<FileName> String type.
The folder where waveform segment file is stored, the folder specified by users can only be
that under relative path.
<WaveForm> String type.
Name of waveform segment file, the maximum number of waveform segment file
supported by this command is 64.
<Reps> Integer. Times of looped playback of each waveform segment.
Range: 1[1
,
65535].
<Marks> Discrete data type, mark switch of each waveform segment file, specific option is in
command format.
Example:
[:SOURce]:RADio:ARB:SEQuence Seq1
,
waveform1
,
12
,
NONE
,
vaveform2
,
300
,
M1M2
Load a waveform sequence, which is located in Seq1 folder which contains
waveform1 and waveform2. The former is played 12 times with no mark output; the
latter is played 300 times, with mark 1 and mark 2 of each symbol output.
Key Entry:
【
Arb
】
—>[Base Config] —>[Add Wave Seg]
Note:
for setting only.
[:SOURce]:RADio:ARB:SEQuence:CLOCk <Mode>
Function:
This command sets Clock Type when the signal generator is in Arb mode. Arb mode
allows users to use CUSTom mode only. Other modes cannot be set; Seq mode
enables users to select CURRent, HIGH or CUSTom. For setting operation mode of
arbitrary waveform, see the command ―
:RADio:ARB:MODE
‖.
Setting format:
[:SOURce]:RADio:ARB:SEQuence:CLOCk CURRent|HIGH|CUSTom
Query format:
[:SOURce]:RADio:ARB:SEQuence:CLOCk?
Parameter:
<Mode> Discrete data. Clock Type in Arb mode is as follows:
CURRent
| 0: Play waveform segment files in Seq mode at sampling rate
of each waveform segment;
HIGH
| 1: Play waveform segment files in Seq mode at maximum
sampling rate of waveform segments each waveform segment;
CUSTom
| 2: Play waveform segment files in Seq mode at the current
clock frequency setting of the signal generator.
Example:
[:SOURce]:RADio:ARB:SEQuence:CLOCk HIGH
Set Clock Type to Highest.
Reset status:
CURR
Key Entry:
【
Arb
】
—>[Base Config] —>[Clock Type]
[:SOURce]:RADio:ARB:SCLock:RATE <ClockRate>
Function:
This command is used for setting the signal sampling rate of arbitrary waveform. The set
Содержание 1465 Series