3 SCPI
3.3 Instrument subsystem command
110
:SYSTem:COMMunicate:LAN:IP ...................................................................... 111
:SYSTem:COMMunicate:LAN:SUBNet ............................................................. 111
:SYSTem:COMMunicate:LAN:GATeway .......................................................... 112
:SYSTem:ERRor[:NEXT] .................................................................................. 112
:SYSTem:PRESet:TYPE ................................................................................... 112
:DIAGnostic:INFormation:CCOunt:PON
Function:
This command is used to query the cumulative number of times the instrument is
powered on
Query format:
:DIAGnostic:INFormation:CCOunt:PON
?
Example:
:DIAGnostic:INFormation:CCOunt:PON
?
This example indicates query of cumulative
startup times of the signal generator.
:DIAGnostic:INFormation:OTIMe
Function:
This command is used to query the instrument firmware date and time stamp
Query format:
:DIAGnostic:INFormation:OTIMe
?
Example:
:DIAGnostic:INFormation:OTIMe
?
This example indicates query of cumulative startup
hours of the signal generator.
:DIAGnostic:SNUM?
Function:
This command is used for reading the system serial number of the signal generator.
Query format:
:DIAGnostic:SNUM
?
Return value:
Serial number
Example:
:DIAGnostic:SNUM This example indicates query of the system serial number of the
signal generator.
Note:
for query only.
:SYSTem:COMMunicate:GPIB:ADDRess <Address>
Function:
This command is used for setting GPIB address of the signal generator. Default address
is 19. To ensure normal communication, local GPIB address shall be different from
other GPIB addresses in the same test system.
Setting format:
:SYSTem:COMMunicate:GPIB:ADDRess <val>
Query format:
:SYSTem:COMMunicate:GPIB:ADDRess?
Parameter:
<Address> Integer data, GPIB address.
Range: 19[0
,
30].
Example:
:SYSTem:COMMunicate:GPIB:ADDRess 19
Set GPIB address of the signal generator to 19.
Reset status:
19
Key Entry:
【
System
】
—>[GPIB Port] —>[Local GPIB Addr]