Parameter Meaning
Description
-u maxPDU
Maximum PDU value in
byte units
* The CXDI automatically
uses 131072 internally for
operation.
* The above value can be
overwritten by designating a
specific value.
* The DICOM standards do not allow
values of 1301073 or higher to be set.
* This parameter is used when the
operator of the connected storage device
requests a size change.
* In DICOM storage devices, note that the
argument title changes to -m.
(
Æ
See the storage device parameters.)
-c callingTitle
Called App Entity Title
* The AE Title setting field is
automatically used in this
setting.
* The above value can be
overwritten by designating a
specific value.
* Note that the meaning is opposite of the
argument -c for DICOM storage
devices.
(
Æ
See the storage device parameters.)
* The entry for the OPU output device
title is used here.
-t callingTitle
Calling App Entity Title
* The CXDI automatically
uses CANON_CCR
internally for the operation.
* The above value can be
overwritten by designating a
specific value.
* Note that the meaning is opposite of the
argument -t for DICOM storage devices.
(
Æ
See the storage device parameters.)
-g
N-GET Printer
compatibility mode
*This parameter is used
simply as “-g”.
* In the CXDI default settings,
the printer information is not
designated. In this case, the
printer side sends all the
information that it has
(DICOM official
specifications).
* When the -g option is added,
the essential information
only is collected. This
information includes the
Printer Status and Printer
Status Info. (To prevent
installation when the printer
does not satisfy the above
DICOM specifications.)
Æ
Normally, this option is not used.
This parameter has been provided as a
remedy when a printer error occurs when
optional devices are not used.
-S
Silent mode
* This parameter is used
simply as “-s”.
* This parameter is used to
prevent the transfer software
from displaying the
debugging character string
on the console.
* This parameter does not affect DICOM
data transfer.
Æ
Silent mode does not need to be
designated since the CXDI automatically
makes the setting internally.
-p
This parameter dumps the
association parameter.
* This parameter is used
simply as “-p”.
* This parameter is used to set
the transfer software so that
the debugging character
string is displayed on the
console.
* This parameter does not affect DICOM
data transfer.
Содержание CXDI-40G COMPACT
Страница 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Страница 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Страница 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Страница 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...
Страница 165: ...CXDI SYSTEM Rev 01 CONTENTS CXDI 40G COMPACT 1 IMAGING UNIT 2 POWER BOX 3 INDEX OF PARTS NUMBERS 4 ...
Страница 167: ...CXDI SYSTEM 1 2 Rev 01 Canon Inc Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Key No 10 Key No 11 ...