2. Installation
- 17 -
<SKINTOP spiral of Sensor cable>
5) Attach the cable mounting plate to the imaging unit. Next, connect the ready lamp connector
and sensor cable connector to the imaging unit connector. After checking that the cables are
routed properly and the length is adjusted properly, secure with the two cable ties.
Note
Allow enough length and route the sensor cable while taking care that excessive force is
not applied to it.
Allow enough length and route the ready lamp cable while taking care that excessive
force is not applied to it.
Gap 1.5 to 2.0 mm
Ready lamp connector
Sensor cable connector