2. Installation
- 9 -
3.2 System Installation Procedure
No.
Step
Conditions and checkpoints
Reference
1
Unpack and check the
product’s constituent parts
- There are no missing parts, damage,
scratches, or other imperfections.
2
Affix the insulation sheets to
the sensor unit
- Affix to the specified locations on the
imaging unit without making any raised or
bent sections.
3
Install the imaging unit
- Install and secure in place in the Bucky unit,
stand or table. Obtain the imaging unit
fixture pieces from a distributor.
- Be careful not to drop the imaging unit.
4
Connect the imaging unit and
power box
- The cables must be routed so that no
unreasonable loads are placed on them.
- R at stationary parts (radius of cable inner
circumference) = 25 mm or more
- R at moving parts (radius of cable inner
circumference) = 50 mm or more
5
Connect the power box and the
control PC
- The cables must be routed in such a way that
no unreasonable loads are brought to bear
upon them.
6
Attach and connect the
operation unit
- The cables must be routed in such a way that
no unreasonable loads are brought to bear
upon them.
7
Adjust the alignment
- Be careful not to scratch the imaging unit.
8
Install the phototimer
- Install in the Bucky unit, stand or table.
- Be careful not to scratch the sensor.
9
Start the system
10
Identifying the imaging units
and setting the number of units
to be connected (inputting the
sensor serial numbers)
“Identifying
the
Sensor Unit” in
“Setting”
11
Install control PC unit serial
number.
“Entering
Control PC
Serial Number”
in “Setting”
12
Check the software program’s
version
- The compatibility of the sensor unit and the
control PC must be checked on the
compatibility list, and the software program
must be installed or upgraded as required.
“Checking the
Firmware
Version” in
“Setting”
13
Check date and time
- Date and time must be changed according to
the area where the instrument is installed.
“Checking and
Setting the Date
and Time”.
14
Connect the power box and
X-ray generators
- The manufacturer of the X-ray generators
must be asked to handle the connections
with the generators.
15
Adjusting the timing with the
X-ray generators
- This is generally not necessary.
16
Perform calibration imaging
- No errors must be displayed.
- Calibration imaging must be performed by
service staff only.
Содержание CXDI-40G COMPACT
Страница 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Страница 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Страница 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Страница 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...
Страница 165: ...CXDI SYSTEM Rev 01 CONTENTS CXDI 40G COMPACT 1 IMAGING UNIT 2 POWER BOX 3 INDEX OF PARTS NUMBERS 4 ...
Страница 167: ...CXDI SYSTEM 1 2 Rev 01 Canon Inc Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Key No 10 Key No 11 ...