3. Function
- 10 -
5. msec maximum value checking
Even if the values recorded in the msec values of calibration/imaging body part within the
BodyPart.ini exceed the maximum msec value of the sensor, the OPU can perform the
exposure using those values without checking. (This function was already implemented
before CXDI v5.1.)
It is the same when using the msec imaging conditions table, if a value bigger than the
sensor’s maximum msec value is written in the BodyPart.ini file, the value will not be
automatically replaced with the value of the imaging conditions table, but can be used for
the exposure as it is.
Note:
In the case of mAs lead, the msec will be automatically recalculated based on the mA,
mAs, and the msec value will be automatically overwritten.
6. kV, mA, mAs, cm maximum value check
If any of the kV, mA, mAs, cm values of the calibration/imaging body part recorded
within the BodyPart.ini file exceeds the maximum value of the sensor, when the CXDI
starts up, a warning message will be displayed and the value will be automatically
modified to be at or below the limit. (If the imaging conditions table is used, a table value
not exceeding the maximum value of the sensor will be specified.)
Содержание CXDI-40G COMPACT
Страница 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Страница 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Страница 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Страница 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...
Страница 165: ...CXDI SYSTEM Rev 01 CONTENTS CXDI 40G COMPACT 1 IMAGING UNIT 2 POWER BOX 3 INDEX OF PARTS NUMBERS 4 ...
Страница 167: ...CXDI SYSTEM 1 2 Rev 01 Canon Inc Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Key No 10 Key No 11 ...