4. Repair Guide
- 8 -
4. Operation Required for Replacing Parts
Perform the following operations when the unit is replaced.
1) Sensor unit
(1) When sensor unit is replaced
1) Set the IP address. (This operation is not required if the default IP address is used.)
Refer to “Imaging unit IP address setting” in the Tool Software Operation Manual.
2) Check the version of firmware and PLD code version.
Refer to “12.2 Checking the Firmware Version” in “12 Settings” in “2. Installation”.
3) Install the firmware and PLD code as required.
Refer to “12.3 Installing Firmware and PLD Code” in “12 Settings” in “2. Installation”.
4) Identify the sensor unit.
Refer to “12.6 Identifying the Sensor Units” in “12 Settings” in “2. Installation”.
5) Change total number of exposures. (Basically the counter should be reset to “0” when the
sensor unit is replaced.)
Refer to “12.13 Changing the Total Image Count” in “12 Settings” in “2. Installation”.
6) Check the image quality.
Refer to “14 Image Quality” in “2. Installation”.
7) Back up the data. (Back up the data to FD as required.)
Refer to “12.14 Backing Up when Installing” and “12.15 Backing Up Setting Data to FD” in
“12 Settings” in “2. Installation”.
(2) When PWB-EDi is replaced
1) Set the PWB dipswitches.
Refer to “3 PCB Settings” of “4. Repair Guide”.
2) Set the IP address. (This operation is not required if the default IP address is used.)
Refer to “Imaging unit IP address setting” in the Tool Software Operation Manual.
3) Check the version of firmware and PLD code version.
Refer to “12.2 Checking the Firmware Version” in “12 Settings” in “2. Installation”.
4) Install the firmware and PLD code as required.
Refer to “12.3 Installing Firmware and PLD Code” in “12 Settings” in “2. Installation”.
5) Install the sensor serial number.
Refer to “12.4 Checking the Sensor Serial Number” in “12 Settings” in “2. Installation”.
6) Install the sensor data file.
Refer to “DP File Write Tool Software” in the Tool Software Operation Manual.
7) Check the image quality.
Refer to “14 Image Quality” in “12 Settings” in “2. Installation”.
Содержание CXDI-40G COMPACT
Страница 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Страница 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Страница 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Страница 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...
Страница 165: ...CXDI SYSTEM Rev 01 CONTENTS CXDI 40G COMPACT 1 IMAGING UNIT 2 POWER BOX 3 INDEX OF PARTS NUMBERS 4 ...
Страница 167: ...CXDI SYSTEM 1 2 Rev 01 Canon Inc Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Key No 10 Key No 11 ...