2. Installation
- 42 -
9.3.2 Network Setting for Screwcap.ini
The CXDI software communicates with the imaging part through screwcap.dll by using the
communication protocol for sending and receiving commands and responses.
In conjunction with the communication, Screwcap.ini retains the information of the connected
imaging unit and control PC such as network addresses.
It is necessary to edit Screwcap.ini to communicate with the imaging part.
Screwcap.ini factory setting is the following default setting (see the figure below
).
If a single
sensor unit to be connected to the control PC, Screwcap.ini does not require editing, and only a
check need be conducted.
If a multiple number (2 or more) of sensor units are to be connected to the control PC,
Screwcap.ini must be edited. (* Refer to “9
.
3.4. Connection setup for a multiple number of
sensor units.”)
Screwcap.ini is located in the following directory:
D:\ccr\screwcap.ini
*1 “SensorNum=1” is the default value for the installer of the system software.
Only the [Sensor1] section is supported.
When the SensorNum key has been set to “2,” “3” or “4” (SensorNum=2, 3 or 4), the
[Sensor1, Sensor2], [Sensor1, Sensor2, Sensor3] or [Sensor1, Sensor2, Sensor3, Sensor4]
sections are supported, respectively.
*2 Imaging unit IP address: This must be identical to the IP address stored in “9.3.3
Network settings stored in imaging unit” in the next section.
Note:
All editing using keys other than the SensorNum keys is forbidden.
[HostInfo]
ResponseTimeout=30000
DataTimeout=60000
IntervalTimeout=30000
[SensorInfo]
SensorNum=1
[Sensor1]
IpAddress=192.168.100.11
CommandPort=12121
DataPort=12122
[Sensor2]
IpAddress=192.168.100.12
CommandPort=12121
DataPort=12122
[Sensor3]
IpAddress=192.168.100.13
CommandPort=12121
DataPort=12122
[Sensor4]
IpAddress=192.168.100.14
CommandPort=12121
DataPort=12122
[Log]
LogLevel=2
[SensorInfo]
SensorNum=1*
¹
In the case of a multiple
number of units, input here
the number of sensors to
be connected.
[Important]
If
“2” has been input for
SensorNum, the IP
address settings of the two
sensor units must match
the IpAddress key in either
the [Sensor1] or [Sensor2]
section without fail.
Imaging unit IP address.
*
²
The factory setting
This is identical to the IP address
written on the Di board inside the
sensor unit.
Содержание CXDI-40G COMPACT
Страница 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Страница 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Страница 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Страница 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...
Страница 165: ...CXDI SYSTEM Rev 01 CONTENTS CXDI 40G COMPACT 1 IMAGING UNIT 2 POWER BOX 3 INDEX OF PARTS NUMBERS 4 ...
Страница 167: ...CXDI SYSTEM 1 2 Rev 01 Canon Inc Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Key No 10 Key No 11 ...