- 68 -
Parameter List (Separate Document 1)
DICOM storage device
In the CXDI, DICOM data transfer is performed using the transfer software module
“send_image” .The settings for these parameters are described below.
Parameter Meaning
Description
-m maxPDU
Maximum PDU value in
byte units
* The CXDI automatically
uses 131072 internally for
operation.
* Designating a specific value
allows overwriting of the
above value.
* The DICOM standards do not allow
values of 1301073 or higher to be set.
* This is used when the operator who
manages the connected storage device
requests a size change.
* In DICOM printing, note that the
argument title changes to -u.
(
Æ
See the printing parameters.)
-t calledTitle
Called App Entity Title
* The AE Title setting field is
automatically applied to this
setting.
* Designating a specific value
allows overwriting of the
above value.
* Note that the meaning is opposite of the
DICOM printing argument -t.
(
Æ
See the printing parameters.)
* The entry for the OPU output device
title is used here.
-c callingtitle
calling App Entity Title
* The CXDI automatically
uses CANON_CCR
internally for the operation.
* Designating a specific value
allows overwriting of the
above value.
* Note that the meaning is opposite of the
DICOM printing argument -c.
(
Æ
See the printing parameters.)
* This is used when the operator who
manages the connected storage device
requests a change in the installed
identification information (version).
-s SOPName
(for reference)
This parameter designates
whether class be connected
for performing association
at the beginning of
transfer.(CR/T/MR/NM/S
C/US)
* This is not used in the
CXDI.
-I
A-RELEASE-RES is
ignored.
* This parameter is used
simply as “-I”
* This is used when the error message
“130012 Peer aborted Association (or
never connected)” occurs even though
the DICOM data transfer was
successful.
Æ
This is used differently based on the
connected storage devices.
-d FAC
This parameter dumps a
specific facility log.
(DCM/DUL/SRV)
* This parameter is used
simply as “-d”
* This parameter is used to
make the transfer software
put the debugging character
string on the console.
* This parameter does not affect DICOM
data transfer.
* This parameter outputs the CXDI log
based on Windows NT.
Содержание CXDI-40G COMPACT
Страница 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Страница 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Страница 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Страница 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...
Страница 165: ...CXDI SYSTEM Rev 01 CONTENTS CXDI 40G COMPACT 1 IMAGING UNIT 2 POWER BOX 3 INDEX OF PARTS NUMBERS 4 ...
Страница 167: ...CXDI SYSTEM 1 2 Rev 01 Canon Inc Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Key No 10 Key No 11 ...