2. Installation
- 15 -
4.3 Attaching the Sensor Cable and Ready Lamp Cable
The sensor cable and ready lamp cable can be attached to the imaging unit either before or after
the imaging unit is installed in the Bucky unit, stand or table. Check the structure of the Bucky
unit, stand or table obtained from the distributor, and attach the sensor cable and ready lamp cable
using the easiest procedure. This manual describes the procedure for attaching the cables before
installing the imaging unit.
Notes
* If the imaging unit is installed to the Bucky unit, stand or table by a single operator
when the sensor cable and ready lamp cable are already connected, there is a risk of the
cables getting pinched or bent. Therefore, be sure to install the imaging unit to the
Bucky unit using two or more operators.
1) Remove the two screws on the connector cover, and remove the connector cover.
2) Remove the two screws on the cable mounting plate, and attach the cable mounting plate from
the imaging unit. Next, remove the two screws on the conductive
holder
, and remove the
conductive
holder
from the cable mounting plate.
Connector cover
Screws (2)
Cable mounting plate
Conductive holder
Содержание CXDI-40G COMPACT
Страница 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Страница 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Страница 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Страница 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...
Страница 165: ...CXDI SYSTEM Rev 01 CONTENTS CXDI 40G COMPACT 1 IMAGING UNIT 2 POWER BOX 3 INDEX OF PARTS NUMBERS 4 ...
Страница 167: ...CXDI SYSTEM 1 2 Rev 01 Canon Inc Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Key No 10 Key No 11 ...