![Canon CXDI-40G COMPACT Скачать руководство пользователя страница 146](http://html1.mh-extra.com/html/canon/cxdi-40g-compact/cxdi-40g-compact_service-manual_4002850146.webp)
3. Function
- 8 -
3 Available Imaging Time (Normal Imaging/Long-term
Imaging)
[Overview]
There are two available imaging times as follow:
1) Normal imaging (0 - 1000 msec)
2) Long-term imaging (1001 - 3000 msec)
The maximum imaging time is now 3 seconds for the CXDI-40G Compact sensor, which can be
set on the GUI.
[Precaution]
•
The preview display becomes slower when using long-term imaging.
•
The preview display may also become slower when performing normal imaging using the
part button that is set to long-term imaging (imaging time sett
ings: 1001 msec to 3000 msec).
•
Imaging preparation time: 10 seconds for normal imaging and 30 seconds for long-term
imaging. In imaging preparation time, the imaging unit changes from sleep status (sensor in
standby status) to ready status (possible to perform imaging).
(The imaging cycle time
*1
becomes longer.)
*1
Time required preparing for the next imaging after a single imaging is done. Normal
imaging: 15 sec, long-term imaging: 30 sec.
[Technical Description]
1. The maximum value for imaging condition
•
When using the CXDI-40G Compact sensor, the maximum value of the exposure time
is 3000 msec, which can be set on the screen.
•
The maximum value of the tube current is 1600mA. The maximum value of the mAs
value is changed from the 3000 maximum value of the exposure time to 4800 mAs.
[Reference]
Exposure time
Tube current
mAs value
Up to CXDI V5.1
1000msec
↓
5000mA
↓
5000mAs
↓
CXDI-11- CXDI-31, 60G sensor
1000msec
1600mA 1600mAs
Other sensors
3000msec 1600mA 4800mAs
Содержание CXDI-40G COMPACT
Страница 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Страница 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Страница 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Страница 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...
Страница 165: ...CXDI SYSTEM Rev 01 CONTENTS CXDI 40G COMPACT 1 IMAGING UNIT 2 POWER BOX 3 INDEX OF PARTS NUMBERS 4 ...
Страница 167: ...CXDI SYSTEM 1 2 Rev 01 Canon Inc Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Key No 10 Key No 11 ...