![Canon CXDI-40G COMPACT Скачать руководство пользователя страница 147](http://html1.mh-extra.com/html/canon/cxdi-40g-compact/cxdi-40g-compact_service-manual_4002850147.webp)
3. Function
- 9 -
2. Imaging condition check at the time of upgrade
If any of the kV, mA, mAs, cm (except msec) values recorded in the imaging condition
table or in each body part information exceeds the limit of the sensor, when the CXDI
starts up, a warning message will be displayed (Fig. 4-1) and the value will be
automatically modified to be at or below the limit. (However, thinking from the range of
values that can be used with the generator, the probability of the warning message is very
low.)
3. Adding values to the imaging conditions table
It is possible to add values up to 3000 msec, 4800 mAs to the imaging conditions table.
However, if the exposure is performed with a CXDI-11 - CXDI-31 or CXDI-60G sensor,
it is not possible to set values higher than 1000 msec, 1600 mAs.
4. Imaging conditions table
It is possible to add values up to 3000 msec and 4800 mAs to the system settings -
imaging conditions table on the settings screen. However, if 1001 - 3000msec, 1601 -
4800mAs values are added, a warning message will be displayed only once, saying the
values cannot be used with the CXDI-11 - CXDI-31 and CXDI-60G sensors. [Figure 3-]
If at the time of a regular exposure or a calibration exposure, the imaging body part of a
CXDI-11 - CXDI-31 or CXDI-60G sensor was selected, values higher than 1000 msec,
1600 mAs cannot be selected from the imaging conditions table.
Figure 3-1
Figure 3-2
Press [OK]
Содержание CXDI-40G COMPACT
Страница 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Страница 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Страница 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Страница 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...
Страница 165: ...CXDI SYSTEM Rev 01 CONTENTS CXDI 40G COMPACT 1 IMAGING UNIT 2 POWER BOX 3 INDEX OF PARTS NUMBERS 4 ...
Страница 167: ...CXDI SYSTEM 1 2 Rev 01 Canon Inc Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Key No 10 Key No 11 ...