2 Installation
- 38 -
8.5 The Connection with X-ray Generator Equipment
+5V(isolated)
Photo coupler
RX_REQ
RX_COM
Relay
PC1
RL1
Connector
J4
POWER BOX
X-ray I/F Cable
X-ray Generator
#2
#1
#4
#3
A2
A1
B2
B1
X-ray permission signal
X-ray request signal
*1:Impedance between A1-A2 < 100ƒ ¶
*2:Permitted Vold AC250Vmax/DC30Vmax
Permitted Current 10mA-2A
WHT
YEL
ORG
GRY
TLP281(GR/F)
NC2D-P-DC5V
Connections Between System and X-ray Generator
Connection conditions
1. The X-ray exposure signal line (including switching mechanism) must be insulated, and its
total impedance must be 100 ohms or less.
2. The maximum contact voltage of the X-ray exposure authorization signal line is AC 250V
and DC 30V, and its current ranges from 10mA to 2A.
Only the insulated secondary power supply can be connected.
3. The protective installation of the X-ray generator must have the same electrical potential as
the system.
Содержание CXDI-40G COMPACT
Страница 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Страница 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Страница 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Страница 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...
Страница 165: ...CXDI SYSTEM Rev 01 CONTENTS CXDI 40G COMPACT 1 IMAGING UNIT 2 POWER BOX 3 INDEX OF PARTS NUMBERS 4 ...
Страница 167: ...CXDI SYSTEM 1 2 Rev 01 Canon Inc Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Key No 10 Key No 11 ...