3. Function
- 6 -
2 Power Box
Main functions
(1) Supply of power to the sensor unit (internal power supply 7m unit)
(2) Interfacing with the X-ray generator
(3) Relay of Ethernet communications while simultaneously providing isolation from the
network to ensure safety (AC 230V, basic insulation)
(4) Provision of switches for starting the sensor unit firmware, initializing codes
The power box consists of the PWB-40EXRAY board (X-ray interface) and power supply,
and it comes with functions to transfer signals to and from the sensor unit, supply power,
and transfer signals to and from the X-ray generator.
Each sensor unit must be connected to one power box and vice versa, and multiple
numbers of units cannot be connected to one box and vice versa.
2.1 PWB-40EXRAY Board
This board comes with the functions described below.
(1) Interfacing with the X-ray generator
(2) Relay of Ethernet communications while simultaneously providing isolation (AC
230V, basic insulation)
(3) Power supply 7m setting (JP1 #6-#8 shorted)
(4) Provision of switches for starting the sensor unit firmware, initializing codes
Control PC
Ready Lamp Unit
Patient environment
Power supply
unit (7m)
AC Power
cable with
plug
Содержание CXDI-40G COMPACT
Страница 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Страница 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Страница 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Страница 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...
Страница 165: ...CXDI SYSTEM Rev 01 CONTENTS CXDI 40G COMPACT 1 IMAGING UNIT 2 POWER BOX 3 INDEX OF PARTS NUMBERS 4 ...
Страница 167: ...CXDI SYSTEM 1 2 Rev 01 Canon Inc Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Key No 10 Key No 11 ...