5.Troubleshooting
- 2 -
1.3.2 When Error Occurs
Information required differs according to whether optional module (such as DMW,
generator communication module, etc.) is used or not.
(1) Log files and information required no matter optional module is used or not
1) How the system is used at the customer’s site (System composition)
2) Version of CXDI host program
3) Error message and code
4) All the INI files directly under CCR folder
5) All the BodyParts folders
All of the above five items are necessary to see how the CXDI is set. If the problem has
occurred in the CXDI, the information is required to have the problem occur again.
6) opu3.log and opu3_YYYYMMDDHHMMSS.log
When file size exceeds 1 MB, opu3.log makes a new “opu3.log” with a different name in
“opu3_YYYYMMDDHHMMSS.log” when the system is restarted.
Ten history files in the past will be saved.
Because a new log is made when the file size is large, log at the time the error occurred
may be in a file other than “opu3.log”.
Since location of the log cannot be checked at the customer’s site, please be sure to collect
not only the opu3.log but all the history files.
7) syslog
8) Dr. Watson log
9) Event viewer system log
10) Sensor log
Other: User information, 5W1H, frequency of occurrence, reproducibility, special
characteristics
(2) Required log files and information according to the used optional module
Please also gather the following files and information other than those indicated in (1).
(2-1) DMW
1) chmwm.ini
2) ccMax_ws.ini
chmwm.ini/ccMax_ws.ini will be collected automatically when all the INI files directly
under CCR folder are collected.
3) chmwm.log
4) DICOM conformance statement of RIS
Содержание CXDI-40G COMPACT
Страница 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Страница 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Страница 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Страница 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...
Страница 165: ...CXDI SYSTEM Rev 01 CONTENTS CXDI 40G COMPACT 1 IMAGING UNIT 2 POWER BOX 3 INDEX OF PARTS NUMBERS 4 ...
Страница 167: ...CXDI SYSTEM 1 2 Rev 01 Canon Inc Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Key No 10 Key No 11 ...