2. Installation
- 2 -
(6)
The patient must not directly touch the imaging unit when taking exposures.
(7)
Do not install on a motor vehicle.
1.3 Installation Notes
(1)
Noise and artifacts tend to appear on the image in environments with strong
electromagnetic fields. Therefore, try to select an installation location that is not near
electronic devices.
Examples of applicable devices:
CRT monitors, X-ray generators, medical electronic devices
(2)
Before starting the installation work, discharge any electrostatic charge that has
accumulated on your body.
(3)
Calibration imaging can be performed by service staff only. It cannot be performed by
users.
(4)
Do not perform calibration imaging with the grid installed.
Except for grid, phototimer and other parts that are located in the X-ray entrance surface must
be installed when calibration imaging is performed.
(5)
After installing the imaging unit in the Bucky unit, stand or table, if the user cannot confirm
the cross lines of the imaging unit surface, attach an indicator mark after adjusting the
X-ray tube position if necessary.
(6)
Routing of the cables for the table, upright stand, and universal stand is based on the table,
upright stand, and universal stand specifications. However, the bending radius R of the
cable (radius of inner cable circumference) is based on the following conditions that
incorporate the bending durability and other factors.
R at stationary parts (radius of cable inner circumference) = 25 mm or more
R at moving parts (radius of cable inner circumference) = 50 mm or more
Do not forcibly bend, fold over, or pull out the cable, and secure the cable so that no
excessive loads are applied to it in any state.
(7)
Install so that there is sufficient space for maintenance and other work. Imaging unit
dimensions: 470 (W)×548 (L)×32 (H) mm
(8)
Do not apply load of over 12kg(118N) to the X-ray entrance surface of the imaging unit
outer cover.
Содержание CXDI-40G COMPACT
Страница 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Страница 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Страница 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Страница 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...
Страница 165: ...CXDI SYSTEM Rev 01 CONTENTS CXDI 40G COMPACT 1 IMAGING UNIT 2 POWER BOX 3 INDEX OF PARTS NUMBERS 4 ...
Страница 167: ...CXDI SYSTEM 1 2 Rev 01 Canon Inc Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Key No 10 Key No 11 ...