5.Troubleshooting
- 4 -
1.3.3 When Image is Bad
* Required Log files
1) How the system is used at the customer’s site (System composition)
2) Version of CXDI host program
3) D:\ccr folder set
4) Trouble image together with the images before and after it, all Bin files and logs folder if
ccr folder set cannot be obtained
5) Extracting sensor log files by "logread.exe"
Refer to SIDR-08-003 for file collection procedure in case of image errors.
1.3.4 Trouble Analysis Tools
This section describes the analysis tools used to remedy trouble or ascertain the status quo.
The tools are useful in analysis applications on-site or at the office.
1) trnsfrm3.exe
* Viewer software for image analysis
* "Technical document for image analysis methods" given in SIDR-03-009 (software must
be purchased separately)
2) Dcap.exe
* Image transfer software
* "Dcap Image Transfer Software User’s Manual" given in SIDR-03-021
3) Pingtest.exe
* Software for evaluating reliability of network communications
* "Operation Verification Results for General-Purpose LAN Card" given in CXDI-50G,
SMR04-009
Содержание CXDI-40G COMPACT
Страница 24: ...2 Installation 4 2 CXDI SYSTEM III Assembly Packaging Accessories Box Power Box ...
Страница 56: ...2 Installation 36 8 3 When Timeout due to RX_REQ not Negating ...
Страница 135: ...2 Installation 115 16 2 Power Box Mass 4 2 Kg not including cable Unit mm ...
Страница 136: ...2 Installation 116 16 3 Ready Lamp Unit Mass 0 5 Kg Unit mm ...
Страница 165: ...CXDI SYSTEM Rev 01 CONTENTS CXDI 40G COMPACT 1 IMAGING UNIT 2 POWER BOX 3 INDEX OF PARTS NUMBERS 4 ...
Страница 167: ...CXDI SYSTEM 1 2 Rev 01 Canon Inc Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Key No 10 Key No 11 ...