Configuring the Mannequin
45
The blue tube connects to the
BLOOD FILL
port and the yellow tube
connects to the
BLOOD VENT
port.
The Battery Compartment and Ports
3. Ensure the yellow pressure relief knob on the trauma fill tank is closed and
pressurize the trauma fill tank by pumping no more than 20 times.
4. Wait until there is good return of fluid in the reclaim bottle (will start
dripping slowly first then accelerate to an almost constant flow; this may
take a few minutes). A constant flow into the reclaim bottle indicates the
blood tank is full.
5. Disconnect the trauma fill tank from the mannequin and depressurize.
The blood tank is ready for use.
External Filling
To fill the blood tank before it is installed in the mannequin:
1. Locate the Blood Tank Adapter accessory.
The Blood Tank Adapter Accessory
The
BLOOD
VENT
port
The
BLOOD
FILL
port
The beige
connector
The red
connector
The white
connector