Care and Maintenance
257
To drain condensation:
1. Locate the Condensation Drain Hose included with the inventory kit.
The Condensation Drain Hose
2. Bring the hose and a small bucket to the simulator location.
3. Bend right knee to lift the thigh to at least a 45-degree angle.
4. Power on the simulator, but do not start Müse. Allow enough time for the
internal compressor to pressurize the system. (The pump will turn off
automatically.)
5. Power off the simulator.
6. Place the tubing end of the Condensation Drain Hose into the small bucket
and then connect the gray fitting onto the simulator’s
EXT AIR
port,
located on the left side of the simulator, with a push and twist motion.
There will be a sudden release of pressure into the bucket. Any
condensation within the system will drain with this exhaust.
The Battery Compartment and Ports
7. Disconnect the Condensation Drain Hose from the simulator.
The
EXT AIR
port