Using Lucina
226
3. Connect the filled syringe to the
IV FILL
port on the left side of the
simulator’s torso.
The Battery Compartment and Ports
4. Depress the syringe until no water remains in the syringe and disconnect
from the
IV FILL
port.
WARNING:
If a flashback does not occur, do NOT inject any fluid and remove the
needle immediately. Repeat the priming directions and ensure you have injected the
needle properly and into the simulated vein.
NOTE:
Fluids and medications can be administered intravenously by attaching the
tubing from a standard IV solution set to the desire IV access site. If administering
more than 300 mL, connect an external empty IV bag and tubing to the
IV DRAIN
port located on the left side of the mannequin’s torso to prevent overfilling the
simulator.
Administering Intrauterine Balloon
The simulator allows for the insertion of an intrauterine balloon used to manage postpartum
hemorrhage.
To insert and fill the intrauterine balloon:
1. Configure mannequin with the postpartum boggy/contracted uterus.
2. Lubricate the end of the balloon with cottonseed oil and insert into the uterus.
3. Fill intrauterine balloon with 500 mL distilled water, per manufacturer’s
recommendations.
To remove the intrauterine balloon, remove the 500 mL water from the balloon and remove the
catheter from the uterus.
The
IV
FILL
port
Содержание Fidelis Lucina
Страница 1: ...User Guide Maternal Fetal Simulator...
Страница 2: ...2015 CAE Healthcare 905K470052 v2 1...
Страница 28: ...Table Of Contents xxvi This page intentionally blank...
Страница 36: ...Equipment Overview 8 This page intentionally blank...
Страница 190: ...Using M se 162 This page intentionally blank...
Страница 226: ...Using the TouchPro CTG Monitor 198 This page intentionally blank...
Страница 266: ...Using Lucina 238 This page intentionally blank...
Страница 298: ...Care and Maintenance 270 This page intentionally blank...
Страница 389: ......