![Cadex C8000 Скачать руководство пользователя страница 66](http://html1.mh-extra.com/html/cadex/c8000/c8000_user-manual_3250378066.webp)
C8000 Battery Testing System User Manual
54
Edit the details as required and press
OK
to save the changes.
See Section
C-code Creation.
Field
Description
C-Code ID
A unique identification name for the C-
Code.
Chemistry
Selects the desired chemistry (Options:
SLA/NiCD/NiMH/Li/LiPH/Primary)
Cells
Enter the number of cells. Number of
cells supported by Battery chemistry:
NiCD and NiMH up to 30 cells in
series (36 Volts nominal).
SLA batteries up to 18 cells in
series (36 Volts nominal).
Li-ion batteries up to 10 cells in
series (36 Volts nominal).
LiPH batteries up to 10 cells in
series (32 Volts nominal).
Primary batteries up to 30 cells in
series (36 Volts nominal.
Discharge only).
Rating
The capacity rating of the cell(s) or
battery pack. Supported Range 0.050-100
(values are stated in Milli-amperage
hour (mAh)).
Chrg Rate
The rate of charge current to be applied.
(Unit (A))