![Cadex C8000 Скачать руководство пользователя страница 225](http://html1.mh-extra.com/html/cadex/c8000/c8000_user-manual_3250378225.webp)
C8000 Battery Testing System User Manual
213
4.
Click the eclipse (“…”) button next to the
Search Program
field to open the Program Editor for Load Capture
programs. All pre-defined programs will appear in a
window and the active program parameters will be
populated in the
Test Parameters
section.
The
Test Parameters
can be modified and will be saved
on selecting “OK”
5.
Click
OK
. The capture begins.
6.
Once the program is loaded, the Channel status reads
LCU
ARMED
until the trigger condition is met. When
capturing, the status reads
LCU CAPTURING
and
FINISH
will appear on completion. A window appears indicating
the status and the Test ID that has been attached to the
captured waveform.
a.
If no trigger condition is present the LCU will
begin capturing immediately. Note: There is an
approximate 3 ~5 second delay between the
program begin initialized and the C8000 begin the
capture. This is normal operation while the Load
Capture Unit is configured for use.
7.
To view the captured waveform, click
Waveforms
in the
Database
menu. The
Waveform Database
window
appears.
8.
Click the
Load Capture
folder and open the corresponding
waveform ID indicated in the pop-up window.