![Cadex C8000 Скачать руководство пользователя страница 291](http://html1.mh-extra.com/html/cadex/c8000/c8000_user-manual_3250378291.webp)
C8000 Battery Testing System User Manual
279
OFFLINE
255
A system operation, such as a firmware
upgrade, is being performed.
CELL MISMATCH
112
Cells are reaching full charge at different
times so the negative slope is not well
defined. The program proceeds to the next
cycle.
Wait until processing is complete.
Battery is old or has been in storage.
Operating time may be reduced. Either
monitor battery use or discard the
battery.
CHARGE
CURRENT LOW
164
Tester cannot provide the requested charge
current
See BAD FUSES OR DRIVER, Code
160.
CHARGE TIMEOUT
144
Charge time has exceeded the expected
value for the battery.
Tester may not have detected the end
of charge conditions.
CHARGING
2
Battery is being charged normally.
RESUMED
32
Power resumed or the battery is re-inserted
into the adapter.
DISCHARGE
CURRENT LOW
162
The tester cannot discharge the battery.
See BAD FUSES OR DRIVER, Code
160.
DISCHARGE
TIMEOUT
142
Battery capacity has exceeded 250% of the
rated capacity.
DISCHARGING
7
Battery is being discharged normally.
Program will continue when cycle has
completed.
HOT BATTERY ON
TRICKLE CHARGE
159
Battery temperature has exceeded its
maximum setting after service has ended
and battery was on trickle charge. The
trickle charge stops.
Lower the room temperature.
Verify that the C-Code is correct for the
battery.
Battery may be in poor condition.
INSERT THE
BATTERY
8
Insert battery in the battery adapter.
LOW VOLTAGE
TIMEOUT1
123
Battery voltage remains low after
approximately one minute into the charge
cycle.
See BATTERY SHORTED, Code 122.
LOW VOLTAGE
TIMEOUT2
124
Battery voltage remains low after
approximately ten minutes into the charge
cycle.
See BATTERY SHORTED, Code 122.
NO BATTERY
10
No battery is inserted, or the inserted battery
is not detected.
Check and clean all contacts.
OFFLINE
255
OVER VOLTAGE
120
Battery voltage is too high.
Discard the battery.
PAUSE
21
PROGRAM IS PAUSED (CUSTOM
PROGRAM FUNCTION 0NLY).
PLATEAU
TIMEOUT
113
Battery was fully charged before full-charge
conditions were met. For SLA and Li
batteries only.
The end-of-charge setting in the C-
code may have to be increased.
Battery may be old and operating time
less than manufacturer's specifications.
PROCESS
COMPLETE
15
Current battery-service program is complete.
PROGRAM HAS
FAILED
16
The program has produced an error. The
Next statement is not processed. The
program has been terminated.
See the detailed display for the exact