![Cadex C8000 Скачать руководство пользователя страница 242](http://html1.mh-extra.com/html/cadex/c8000/c8000_user-manual_3250378242.webp)
C8000 Battery Testing System User Manual
230
5.
In the
Waveform Name
field, type a name for this
waveform.
6.
In the
Comment 1
and
Comment 2
fields, type any notes
about this waveform.
7.
Click
OK
. The new waveform appears in the
Waveform
Database
window.
3.12.2.1
Add a Step to a Waveform
Note: A zero mA step is required when transitioning between
charge/discharge and vice-versa. The last waveform step must
respect this rule when it is compared with the first waveform step.
1.
Click
Add
in the
Waveform Editor
window. The new step
appears in the
Custom Waveform
window.
2.
Change the values in the
Duration
and the
Charge/Discharge
Current
fields, as needed.
3.12.2.2
Remove a Step from a Waveform
1.
Select the step to delete.
2.
Click
Delete
.
3.12.2.3
Copy a Step to Another Step in a Waveform
1.
Select the step to copy.
2.
Click
Copy
.