Test Type
Sub-Parameter
Description of the Test Type Function
(Default: "0")
SM-Bus Voltage
Requires Voltage value parameter
(V) (Default "0.000v)
Compares the Voltage level of a SM-bus register (0x15)
against the Voltage level enter in the step. Comparison is
conducted every 200mS, 5 seconds after the start of the
program step.
SM-Bus Current
Requires Current value parameter
(A) (Default "0.000A)
Compares the Current reading of a SM-bus register
(0x14) against the current level enter in the step.
Comparison is conducted every 200mS, 5 seconds after
the start of the program step.
Once a
Test Type
is selected the action will need to be
programmed with the comparator function to determine a “PASS”
or “FAIL”. The available comparators available are:
Comparator Table
Comparator
Description of the Comparator Function
<
Less than
<=
Less then or equal to
=
Equal to
>=
Greater than or equal to
>
Greater than
<>
Not Equal to
The selected
Test Type
with the selected
comparator
can be
viewed in equation format in the main window under the “Test”
column.
The following actions are available for selection based on a
“TRUE” or “FALSE” action.
TRUE/FALSE action table
True/False Action
Required Parameter
Description of the Program Parameter
Goto Next Step
None
Automatically proceed to the next step. If the current
step is the last step in the program, this action will return
to step zero (0).
Goto Step xx (1-99)
Step number
Automatically proceeds to the programmed step, in the
parameter. This action can not goto step zero (0)
Resets loop counter, goto next
step
None
Resets the counter to zero (0) and will proceed to the
next step. If the current step is the last step in the
program, this action will return to step zero (0).
Inc loop counter, goto xx (1-99)
Step Number
Automatically increments the internal counter by one (1)
and proceeds to the programmed step, in the parameter.