C8000 Battery Testing System User Manual
231
3.12.2.4
Move the Order of Steps in a Waveform
To move the position of a step up or down in the waveform, click
the field name and then click
Move Up
or
Move Down
. This sets
the order that the steps will be executed in the waveform.
3.12.3
Organize Waveforms in Folders
Waveforms can be organized, or grouped, into folders. The
Waveforms database is divided into three groups. A sub-group
can be created within each of these three groups. The following
table describes the Waveform groups:
Group Name
Description
Custom
All standard supported waveforms. Two types
of waveforms are categorized as custom
waveforms:
Waveform #1: Maximum 50 steps with
charge/discharge steps of any duration.
(Minimum 500s, Maximum 12hrs)
Waveform #2: Maximum 500 steps with
charge/discharge steps with a single duration
for each step. (Minimum 500us, Maximum
12hrs)
Load Capture
Displays all captured waveforms from the
accessory Load Capture Unit. Waveforms
cannot be manually added to this group. All
other functions are available.
Playback
Discharge waveforms supported only. No
charge pluses are allowed within this group.
Only one type of waveform is categorized as
playback waveforms:
Waveform #1: Maximum 500 steps, with
discharge steps with a single duration for each
step. (Minimum 500us, Maximum 12hrs)
To move a waveform to a folder:
6.
From the
Databas
e menu, click
Waveforms
. The
Waveform Database
window appears.