![Cadex C8000 Скачать руководство пользователя страница 280](http://html1.mh-extra.com/html/cadex/c8000/c8000_user-manual_3250378280.webp)
C8000 Battery Testing System User Manual
268
Field
Description
Compatible
Chemistry
Limits
Units
Additional
requirements
Typical Range(s)
SM-BUS
Imported
Registers
Min.
Max.
Notes:
Rating>100ah
0.200
SLA, Li,
LiPh
Rating<=100ah
0.055
Rating>100ah
0.200
Discharge
Current
Discharge rate
All
Rating<=100ah
0.055
=
Maximum
Discharge
A
----
0.01~2C
0x14
----
Rating>100ah
0.200
Low
temperature
Low
temperature
limit
All
-30
80
°C
<
Maximum
Temperature
0 ~ 10
No
Zero value is
valid
High
temperature
High
temperature
limit
All
-10
<(Maximum
Temperature -
5)
°C
----
40 ~ 50
No
Zero value is
valid
End of
discharge
End of
discharge
Voltage
All
>
Minimal Voltage
(NiCD, NiMH)
<
Maximum
Voltage
V
----
(NiCD, NiMH)
0.76~1.12V*cell
No
----
(SLA)
1.36~1.95V*cell
(SLA, Li, LiPh)
<
Charge
Voltage
(Li)
2.30~ 3.20V*cell
(LiPh)
1.90~2.35V*cell