![Cadex C8000 Скачать руководство пользователя страница 49](http://html1.mh-extra.com/html/cadex/c8000/c8000_user-manual_3250378049.webp)
C8000 Battery Testing System User Manual
37
output port information (2.7.1.4)
The following table describes the fields shown in the Channel
screen.
Field
Description
Battery ID
The battery’s identification. This field is
blank of a Battery ID has not been added to
BatteryLab.
Program
The name of the program that is being run.
Cycles
Indicates the number of charge and
discharge cycles that have been run during
any given program.
Test Results
Indicates the file name to which the current
program’s results will be saved to, during
operation. On completion of a program,
indicates the file to which the results have
been saved. Note: BatLab will generate an
impendent TEST ID and will incorporate the
data from the C8000 test result file.)
C-Code
The battery parameters being applied by the
C8000.
Step
The step number for the action currently
executing within the program.
Volts
Across the battery terminals. The current
Voltage across battery sense terminal lines.
(Screen is updated once per second.)
Amps
Charge or discharge current. (Screen is