C8000 Battery Testing System User Manual
192
available the Ports must be configured as Output. (See
Program
Parameter Table
)
The Digital Outputs can be set as either high, by placing a check
mark in the box beside the corresponding Digital Output port and
set low (Default) by clearing the check mark.
The Analog Output can be set to an output a voltage level
changing the percentage value for the corresponding Analog
Output Port. A value of 0% will output zero (0) volts while a value
of 100% will output a voltage of five (5) volts.
Each step with in a program can be assigned a unique name, for
identification purposes, by changing or adding text to the
Step
Description
field.
Once all the steps of the program have been added the
information for the program may be entered.
Click on the
Program Parameter
tab and enter information in to
the fields. The fields’ functionalities are defined as shown in the
following table.
Program Parameter Table
Field
Default State
Description of the Program Parameter
Program Name
Blank
The name of the program (This is a required field)
Active
Checked
Allows the program to be used in a service.
Removing the check will remove this program from the
available list when attempting to service a battery.
Smart Battery
Unchecked
Enables the program to be used with SM-BUS C-codes
Term. Data Sample Rate
Default, 500 times
per second (2ms)
The C8000 will automatically save the last 500 data
points for the Voltage and Current at a high capture rate.
This parameter will program the rate in which the C8000
sample the last 500 points.
Options available:
Default, 500 times per second (2ms)
100 times per second (10ms)
50 times per second (20ms)
25 times per second (40ms)
20 times per second (50ms)
10 times per second (100ms)
5 times per second (200ms)
Comment 1
Blank
Comment Field that can be used to store information