C8000 Battery Testing System User Manual
229
recognize the type of waveform that is being created and the
intended device where the waveform is to be used.
Add New Waveform
Device
Compatibility
The Device Compatibility option selects the
device that the waveform is intended for use.
Options available:
C8000: supports both charge and discharge
steps at 500us intervals. (48V/10A/100W)
BK Precision supports only discharge steps at
1ms intervals.
Models Supported:
8510 – (120V/120A/600W)
8514 – (120V/240A/1200W)
8518 – (60V/240A/1200W)
8520 – (120V/240A/2400W)
Note:
If there is a charge step in a waveform
and a BK device has been selected a five
second “0” current step is required between
the charge and discharge currents. All charge
steps will be performed by the C8000.
Maximum Steps This option sets the maximum number of steps
that a waveform will support.
50 steps with charge/discharge currents and
multiple step durations.
(Minimum 500us, Maximum 12hrs per step)
500 steps with charge/discharge currents and
only a single step duration value for all steps.
(Minimum 500us, Maximum 12hrs)
3.
Click the
DOWN ARROW
in the
Device Compatibility
field and
select the appropriate entry.
4.
Select the
Maximum Steps
for the new waveform. Click
OK
.
The
Waveform Editor
window appears.