C8000 Battery Testing System User Manual
274
12
COLD WAIT
THE BATTERY TEMPERATURE BEING REPORTED VIA THE
CHANNEL THERMISTOR IS BELOW THE “LOW
TEMPERATURE LIMIT” (C-CODE) AND WAITING FOR THE
TEMPERATURE TO RISE 5°C ABOVE, BEFORE CONTINUING
SERVICING THE BATTERY.
13
HOT WAIT
THE BATTERY TEMPERATURE BEING REPORTED VIA THE
CHANNEL THERMISTOR IS BELOW THE “HIGH
TEMPERATURE LIMIT” (C-CODE) AND WAITING FOR THE
TEMPERATURE TO FALL 5°C BELOW, BEFORE CONTINUING
SERVICING THE BATTERY
14
OVERHEAT
THE BATTERY TEMPERATURE BEING REPORTED VIA THE
CHANNEL THERMISTOR IS ABOVE THE “MAX
TEMPERATURE LIMIT” (C-CODE. PROGRAM HAS
TERMINATED
15
FINISHED
BATTERY SERVICE WAS COMPLETED SUCCESSFULLY AND
CHANNEL HAS EITHER ENTERED STANDYBY OR HAS
TERMIANTED THE SERVICE, AS DEFINED BY THE
PROGRAM (AS DEFINED BY STEP “0” IN A PROGRAM)
16
PROGRAM FAIL
BATTERY SERVICE HAS FAILED
17
REMOVED
THE BATTERY HAS BEEN DISCONNECTED FROM THE
BATTERY TERMINALS
18
RESERVED
RESERVED
19
RESTING
NO CURRENT IS ACTIVELY BEING APPLIED TO THE
BATTERY THROUGH THE BATTERY PORTS
20
INSERTED
A BATTERY IS CONNECTED TO THE CHANNEL AND THE
CHANNEL SCREEN IS ACTIVE AWAITING BATTERY
PARAMETERS (C-Code) AND PROGRAM TO BEGIN
SERVICING
21
PAUSE
THE BATTERY SERVICE HAS BEEN SUSPENDED AS
DEFINED BY THE PROGRAM
22
RESERVED
RESERVED
23
RESERVED
RESERVED
25
RESUME
CYCLE RESUMED
26
SUSPENDED
THE BATTERY SERVICING HAS BEEN PLACED IN
“SUSPEND” BY THE USER
27
OHMTEST
A DC RESISTANCE TEST IS BEING CONDUCTED
28
IMPEDANCE
AN AC RESISTANCE TEST IS BEING CONDUCTED
29
LCU ARMED
LOAD CAPTURE DEVICE IS ARMED
30
CYCLE DONE
THE CURRENT SERVICE CYCLE HAS BEEN COMPLETED
31
DISCHARGE DONE
THE CURRENT DISCHARGE SERVCICE HAS BEEN
COMPLETED
32
RESUMED
THE BATTERY SERVICE HAS RESUMED AS DEFINED BY
THE PROGRAM AFTER BEING EITHER PLACED IN
”SUSPEND” BY THE USER OR PLACED IN “PAUSE” BY THE
PROGRAM
33
USER TIMEOUT
THE SERVICE HAS TIMED-OUT AS DEFINED IN THE C-CODE
34
BATTERY LAB
SERVICE IS WAITING FOR COMMUNICATION WITH
BATTERYLAB SOFTWARE