C8000 Battery Testing System User Manual
188
Action
Required Action Sub-
parameters
Description of the Action
chambers
Parallel Discharge
Type
:
Allows the selection of the
type of discharge to perform.
Current:
Discharges at a specified
current. (The setting of
”Default” will automatically
double the Discharge Rate
specified in the C-code.)
C-Rate:
Discharges at the rate
indicated.
Watt:
Discharges at a constant
power.
Waveform:
Performs a discharge as
specified by the waveform
parameters.
Custom –Absolute:
Runs the Waveform as
programmed
Accessory Requirement:
See section
Error! Reference source not found.
Error!
Reference source not found.
Example of Current Rate set to Default:
Standard Discharge: 10A/80W
Parallel Discharge: 20A/160W
Within each action step the program has the ability to run a test
during the action step and take appropriate action based on the
result of the test. The table below lists the
Test Type
available for
each action step.
Test Types Table
Test Type
Sub-Parameter
Description of the Test Type Function
No Test (Default)
None
No additional test is performed
Capacity to
Target
None
Compares the latest calculated Capacity value for the
service against the "Target Capacity" value in a C-code.
Note: If the service "Capacity Value" is "N/A" the test
will not be performed. (Test is performed at the end of
the step action)
Capacity diff
Requires capacity value.
(Percentage) (Default: "0%")
Compares the latest calculated Capacity value for the
service against the previously calculated Capacity value
within the same program. Note: 1) A minimal of two
"Discharge" actions are required within the same
program. 2) If the service "Capacity Value" is "N/A" the
test will not be performed. (Test is performed at the end