C8000 Battery Testing System User Manual
185
Action Parameter Tables
Available Actions
Action
Required Action Sub-
parameters
Description of the Action
Finished - No action
None
On completion of service, stop all monitoring. Available
for Step zero (0) only
Finished – Standby
None
On completion of service, monitor the battery. Take
action when monitor conditions are met. NIXX –
perform trickle charge as per C-code.
SLA – perform float charge will occur when the battery
voltage drops below the Standby voltage parameter in
the C-code. The current applied shall equal the current
termination point (CTP) in the C-code.
All Chemistries – Recharges the battery if the voltage
level drops below the standby voltage of the c-code.
Available for Step zero (0) only
Discharge
Type
:
Allows the selection of the
type of discharge to perform.
Current:
Discharges at a specified
current. Specify the current.
C-Rate:
Discharges at the rate
indicated.
Watt:
Discharges at a constant
power.
Waveform:
Performs a discharge as
specified by the waveform
parameters.
Custom –Scaled:
Scales the waveforms
greatest amplitude pulse to
the Discharge Rate in the C-
C-code and if required further
scales the waveform to be
within the C8000 power
requirements.
Custom –Absolute:
Runs the Waveform as
Performs a Discharge on the battery.
Default:
shall use the discharge current value in the C-
code to set the rate of discharge.