![Cadex C8000 Скачать руководство пользователя страница 279](http://html1.mh-extra.com/html/cadex/c8000/c8000_user-manual_3250378279.webp)
C8000 Battery Testing System User Manual
267
Field
Description
Compatible
Chemistry
Limits
Units
Additional
requirements
Typical Range(s)
SM-BUS
Imported
Registers
Min.
Max.
Notes:
Voltage
LiPh
0.750 *
cells
----
Primary
0.320 *
cells
----
Maximum
Safe
Operating
Voltage
The maximum
Voltage allowed
for the battery
NiMH &
NiCD
----
1.800 *
cells
V
45.000
----
No
Primary
Chemistry has
two Maximum
voltages. The
selection
depends on the
Nominal
voltage
selected by the
user.
SLA
2.800 *
cells
48.048
Li
4.500 *
cells
LiPh
4.200 *
cells
Primary (1.2v
nominal)
1.800 *
cells
Primary (3v
nominal)
4.500 *
cells
Maximum
Charge
Maximum safe
operating
charge current
NiMH &
NiCD
0.030
10.000
A
----
----
No
----
SLA, Li,
LiPh
0.055
Maximum
Discharge
Maximum safe
operating
discharge
current
All
Rating<=100ah
0.030
10.000
A
----
----
No
----
Rating>100ah
0.200
Maximum
Temperature
Maximum safe
operating
temperature
All
----
105
°C
----
----
No
----
Charge
Current
Charge rate
NiMH &
NiCD
Rating<=100ah
0.030
= Maximum
Charge
A
----
0.01~2C
0x14