11 Technical specifications
82
ARC 300 / ARC 350 Operating manual
900-350_IFU_V2.6_10296-S13-20100111-EN
Setting
Program
DryCUT
12
ARC350DC_SE_080124_1
0
50
100
150
200
250
10
100
1000
10000
P [W]
DryCut 100W
Blend 0-9
DryCut 200W
Blend 0-9
Measurement of resistances
•
Diagram of power output P [W] as a
function of the load resistance R [
Ω
] for
the setting “DryCUT” = 200 W/100 W
ARC350DC_SE_080125_2
0
20
40
60
80
100
140
120
180
160
200
0
20
40
60
80
100
120
140
160
180
200
displayed value
P [W]
•
Diagram of power output P [W] as a
function of the "DryCUT" setting.
Measured load resistance = 500
Ω
Blend 0-9
ARC350DC_SE_080124_3
0
200
600
1000
1400
400
800
1200
1600
1800
0
20
40
60
80
120
100
140
160
180
200
displayed value
U [Vp]
•
Diagram of HF output voltage U [Vp] as
a function of the "DryCUT" setting (idle
mode)