2 Safety
22
ARC 300 / ARC 350 Operating manual
900-350_IFU_V2.6_10296-S13-20100111-EN
2.5.1. Surgical
environment:
prevention of explosions/combustion
Sparks fly during proper use of the HF device!
X
Do not use the HF device in areas where there is a risk of
explosion.
X
Do not use any flammable or explosive liquids.
X
If display components fail, do not use the HF device any longer!
X
Avoid using ignitable anesthetics and gases which support
combustion (e.g. nitrous oxide, oxygen) during operations (e.g. in
the head and thoracic regions)
.
X
Wear suitable gloves during operations.
X
Use exclusively non-flammable cleaning agents, disinfectants and
solvents (for adhesives). If you use flammable cleaning agents,
disinfectants and solvents, ensure that these substances have
evaporated before using the HF surgery device.
X
Ensure that no flammable liquids collect beneath the patient or in
bodily cavities (e.g. the vagina). Suction and/or flush bodily cavities
before activating the device.
X
Wipe off all liquids before using the HF device.
X
Ensure that no endogenous gases are pesent that could ignite.
X
Ensure that all materials saturated with oxygen (e.g. cotton, gauze)
are removed far enough from the HF environment that they cannot
ignite.
Содержание ARC 300
Страница 1: ...OPERATING MANUAL HIGH FREQUENCY SURGICAL DEVICE...
Страница 2: ......
Страница 4: ...4 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...
Страница 5: ...900 350_IFU_V2 6_10296 S13 20100111 EN ARC 300 ARC 350 Operating manual 5...
Страница 8: ...Legend 8 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...