11 Technical specifications
900-350_IFU_V2.6_10296-S13-20100111-EN
ARC 300 / ARC 350 Operating manual
87
Setting
Program
Moderate COAG Standard
0-6, 9, 12, 14-16, 17,
11 (31 W and above),
23 (only socket 1)
24-99 (socket 1+2)
ARC350MBCOAG_SE_080124_1
0
20
40
60
80
100
120
140
10
100
1000
10000
P[W]
P=60W
P=120W
Measurement of resistances
•
Diagram of power output P [W] as a
function of the load resistance R [
Ω
] for
the setting “Moderate COAG Standard”
= 120 W/60 W
ARC350MBCOAG_SE_080124_2
0
20
40
60
80
100
120
0
20
40
60
80
100
120
displayed value
P [W]
•
Diagram of power output P [W] as a
function of the “Moderate COAG
Standard” setting. Measured load
resistance = 75
Ω
ARC350MBCOAG_SE_080124_3
0
20
40
60
80
100
120
140
160
180
200
0
20
40
60
80
100
120
displayed value
U [Vp]
•
Diagram of HF output voltage U [Vp] as
a function of the “Moderate COAG
Standard” setting (idle mode)