11 Technical specifications
900-350_IFU_V2.6_10296-S13-20100111-EN
ARC 300 / ARC 350 Operating manual
107
Setting
Program
Bipolar LIGATION
2, 3, 4, 16 (only socket 3)
ARC350BipLIG_SE_100111_1
0
50
100
150
200
250
300
10
100
1000
R [Ohm]
P [W]
Measurement of resistances
•
Diagram of the power output P [W] as a
function of the load resistance R [
Ω
] for
the setting “Bipolar LIGATION” “Effect
1-4”
ARC350BipLIG_SE_100111_2
0
50
100
150
200
250
300
4
3
2
1
Effekte
P [W]
•
Diagram of power output P [W] as a
function of the "Bipolar LIGATION"
setting. Measured load resistance =
25
Ω
•
Diagram of HF output voltage U [Vp] as
a function of the “Bipolar LIGATION”
setting (idle mode)