11 Technical specifications
94
ARC 300 / ARC 350 Operating manual
900-350_IFU_V2.6_10296-S13-20100111-EN
Setting
Program
Forced COAG Micro Mixed, Micro Non-cutting,
Micro Cutting
7, 10,
11 (up to 30 W)
ARC350MFCOAG_SE_080124_1
0
5
10
15
20
25
30
35
10
100
1000
10000
P [W]
Micro Forced
Coag P=15W
Micro Forced
Coag P=30W
Measurement of resistances
•
Diagram of power output P [W] as a
function of the load resistance R [
Ω
] for
“Forced COAG Micro Mixed, Forced
COAG Micro Non-cutting, Forced
COAG Micro Cutting” = 30 W/15 W
ARC350MFCOAG_SE_080124_2
0
5
10
15
20
25
30
0
5
10
15
20
25
30
displayed value
P [W]
•
Diagram of power output P [W] as a
function of the setting “Forced COAG
Micro Mixed, Forced COAG Micro Non-
cutting, Forced COAG Micro Cutting”.
Measured load resistance = 500
Ω
ARC350MFCOAG_SE_080124_3
0
500
1000
1500
2000
2500
3000
0
5
10
15
20
25
30
displayed value
U [Vp]
•
Diagram of HF output voltage U [Vp] as
a function of the “Forced COAG Micro
Mixed” setting (idle mode)
ARC350MFCOAG_SE_080124_5
0
500
1000
1500
2000
2500
3000
3500
4000
0
5
10
15
20
25
30
displayed value
U [Vp]
•
Diagram of HF output voltage U [Vp] as
a function of the “Forced COAG Micro
Non-cutting” setting (idle mode)
ARC350MFCOAG_SE_080124_4
0
200
400
600
800
1000
1200
0
5
10
15
20
25
30
displayed value
U [Vp]
•
Diagram of HF output voltage U [Vp] as
a function of the “Forced COAG Micro
Cutting” setting (idle mode)
Содержание ARC 300
Страница 1: ...OPERATING MANUAL HIGH FREQUENCY SURGICAL DEVICE...
Страница 2: ......
Страница 4: ...4 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...
Страница 5: ...900 350_IFU_V2 6_10296 S13 20100111 EN ARC 300 ARC 350 Operating manual 5...
Страница 8: ...Legend 8 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...