11 Technical specifications
900-350_IFU_V2.6_10296-S13-20100111-EN
ARC 300 / ARC 350 Operating manual
95
Setting
Program
GastroCUT ForcedCOAG
21, 22 (only Socket 2)
ARC350GCFC_SE_080125_1
0
10
20
30
40
50
60
10
100
1000
10000
P [W]
P=25W
P=50W
Measurement of resistances
•
Diagram of power output P [W] as a
function of the load resistance R [
Ω
] for
the setting “GastroCUT ForcedCOAG”
= 50 W/25 W
ARC350GCFC_SE_080125_3
0
5
15
25
35
45
10
20
30
40
50
0
5
10
15
20
30
25
35
40
45
50
displayed value
P [W]
•
Diagram of power output P [W] as a
function of the “GastroCUT
ForcedCOAG” setting. Measured load
resistance = 500
Ω
ARC350GCFC_SE_080125_2
0
500
1000
1500
2000
2500
3000
3500
4000
0
5
10
15
20
25
30
35
40
45
50
displayed value
U [Vp]
•
Diagram of HF output voltage U [Vp] as
a function of the “GastroCUT
ForcedCOAG” setting (idle mode)