5. Preparation
Pos: 30 /679-BOWA/ARC 300/350/6 Vorbereitung/6.1 HF-Gerät aufstellen @ 4\mod_1261040578397_6.doc @ 34938 @ 2
5.1.
Setting up the HF device
NOTE
When the HF device is used as intended, electromagnetic fields
are generated! This can affect other devices.
X
Ensure that no electronic devices are set up in the vicinity of the
HF device.
HF devices may be used only in rooms used for medical purposes that
meet the requirements of DIN VDE 0100-710.
If the HF device was previously stored or transported at temperatures
below +10 °C, it will take approximately three hours to adjust to room
temperature.
1. Place the HF device on one of the following platforms:
•
a table,
•
an equipment trolley,
•
a console suspended from a ceiling support or wall-
mounted brackets.
2. Set up the HF device with sufficient clearance to other electronic
devices.
3. Position the HF device with the front of the device facing the
patient/surgeon.
4. Except for the additional device ARC PLUS, do not place any other
devices on the HF device.
5. Do not place any other objects on / above the HF device.
6. Do not set up the HF device on other devices.
7. Connect the power cable.
Pos: 31 /679-BOWA/ARC 300/350/6 Vorbereitung/6.2 HF-Gerät einschalten @ 4\mod_1261040583319_6.doc @ 34957 @ 2
Содержание ARC 300
Страница 1: ...OPERATING MANUAL HIGH FREQUENCY SURGICAL DEVICE...
Страница 2: ......
Страница 4: ...4 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...
Страница 5: ...900 350_IFU_V2 6_10296 S13 20100111 EN ARC 300 ARC 350 Operating manual 5...
Страница 8: ...Legend 8 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...