6 Operation
42
ARC 300 / ARC 350 Operating manual
900-350_IFU_V2.6_10296-S13-20100111-EN
You can change the following factory settings:
Select Cutting settings
1. Press the key/display for „Monopolar Output 1“
35
or „Monopolar
Output 2“
39
and select connection socket 1 or 2.
2. Press the power limitation key for monopolar cutting
3
or
4
.
–Or–
1. Press the key/indicator for "Bipolar Output 3"
44
or "Bipolar
Output 4"
48
.
2. Press the power limitation key for bipolar cutting
23
or
24
.
Set blend factor
1. Press the key for the monopolar degree of scabbing
5/6
.
–or–
Press the key for the bipolar degree of scabbing
25/26
.
2. Set the blend factor between "1-9" or to "0" for coagulation-free
cutting.
Select current type for coagulation current
1. Press the key/indicator for "Moderate Coagulation" current
12
– or –
Press the key/indicator for "Forced Coagulation" current
13
– or –
Press the key/indicator for "Spray Coagulation" current
14
.
2. Set the power limit for the monopolar coagulation current using the
keys for power limitation for monopolar coagulation
10/11
.
–Or–
1. Set the power limit for the bipolar coagulation current using the
keys for power limitation for bipolar coagulation
30/31
.
Pos: 38 /679-BOWA/ARC 300/350/7 Bedienung/7.3 Programm-Beschreibungen @ 4\mod_1261040633397_6.doc @ 35090 @ 2333333333333333333333333
Содержание ARC 300
Страница 1: ...OPERATING MANUAL HIGH FREQUENCY SURGICAL DEVICE...
Страница 2: ......
Страница 4: ...4 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...
Страница 5: ...900 350_IFU_V2 6_10296 S13 20100111 EN ARC 300 ARC 350 Operating manual 5...
Страница 8: ...Legend 8 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...